电感耦合等离子体原子发射光谱法检测磷酸中硫酸根含量的不确定度讨论
Discussion on uncertainty of determination of sulfate content in phosphoric acid by ICP-AES method
摘要
为精准评估电感耦合等离子体原子发射光谱(ICP-AES)法测定湿法磷酸中硫酸根含量的检测可靠性,明确检测结果的关键影响因素,梳理实验全过程,从试样称量、硫酸根标准溶液配制、标准曲线拟合查算硫酸根浓度、检测结果重复性4个维度拆分并量化各不确定度分量,计算合成相对标准不确定度与扩展不确定度.实验结果表明:该方法测定湿法磷酸中硫酸根的合成相对标准不确定度为1.11%,置信水平95%、包含因子k=2时,扩展相对不确定度为2.22%,检测结果可表示为3.20%±0.07%;其中标准曲线拟合查算硫酸根浓度是影响检测结果准确性的核心关键因素,而试样称量、检测重复性引入的不确定度影响极小.
To evaluate the detection reliability of the inductively coupled plasma atomic emission spectrometry(ICP-AES)method for the determination of sulfate content in wet-process phosphoric acid and identify the key factors affecting the detection results,the uncertainty components are split and quantified from four dimensions:sample weighing,sulfate standard solution preparation,concentration calculation of sulfate via standard curve fitting,and detection result repeatability,and the combined relative standard uncertainty and expanded uncertainty are calculated.The results show that the combined relative standard uncertainty of sulfate in wet-process phosphoric acid determined by this method is 1.11%.At a 95%confidence level with a coverage factor of k=2,the expanded relative uncertainty is 2.22%,and the detection result is expressed as 3.20%±0.07%.The concentration calculation of sulfate via standard curve fitting is the core key factor affecting the accuracy of detection results,while the uncertainty introduced by sample weighing and detection repeatability have negligible influence.
年,卷(期):
2026, 41(9)
分类号:
TQ075
机标分类号:
O657.31
TG115.33
R284.1
在线出版日期:
2026年10月12日